01
New adjustment aid
Full control of the firing angle during the procedure, with 365-degree rotation.
Lateral emitting probe
For more power, multiple wavelengths, operator precision and increased durability
SideFire by NeuLight® is an improved, high-powered fiber optic delivery device suitable for laser driven prostate enucleation procedures. It emits laser energy at a 90-degree angle from the axis of the fiber.
Re-imagined for the urologist
The capillary tip is attached to the body of the probe using a proprietary process that maximises energy transmission. The aiming grip is ergonomically designed, allowing for 365-degree rotation and increased durability. The packaging is damage-proof and easy to use, so each sterile device can be seamlessly passed off by the scrub technician to the surgeon.
Our NeuLight® branded lateral emitting fiber probe exceeds OEM laser system power requirements and is compatible within varying continuous treatment wavelengths such as thulium. Our SideFire is cleared by the FDA and suited for high-power medical laser applications.
New design
01
Full control of the firing angle during the procedure, with 365-degree rotation.
02
Interfaces with open-port laser systems already in the theatre.
03
Allows for easy, single-handed removal of the product.
04
Protects the distal leg and lets the scrub technician pass the device off seamlessly.
| Max. input power | 200 W continuous wave for 2 minutes 100 W continuous wave for 30 minutes |
|---|---|
| Core diameter [µm] | 600 |
| Clad diameter [µm] | 660 |
| Numerical aperture | 0.22 |
| Length | 3 m |
| Connector type | SMA905 |
| Wavelength | approx. 2000 nm in the NIR spectral range Holmium wavelength is not validated for use |
| Usability | Sterile, ready to use with open port lasers, disposable |
| NeuLight® product code | WFS-600-SF |
| Custom design | Private label possible Integration of RFID technology possible |
Source: SideFire by NeuLight®, 07.2024/en. Subject to technical changes without notice.
Clinical context
A typical application for SideFire by NeuLight® is the therapeutic treatment of Benign Prostatic Hyperplasia (BPH), a common condition in aging men that affects approximately 50 million people in the United States.
An enlarged prostate generally results in a variety of urologic symptoms such as blockage of urine flow, kidney stones, bladder stones and urinary tract issues.
We will confirm compatibility with your system’s wavelength and power profile, and discuss private-label or RFID-enabled versions.